myLabComplianceCLIA compliance software · lab-consultants.comBook a demo

CAP survey preparation: a practical guide

How CAP-accredited and applicant laboratories can stay ready between inspections, based on CAP's published checklists and guidance.

Book a demo Visit mylabcompliance.io

How CAP inspections work

The College of American Pathologists (CAP) Laboratory Accreditation Program is a CMS-approved accreditation option for CLIA. Accredited labs are inspected on a two-year cycle by a peer inspection team, and the inspection is assessed against the CAP Accreditation Checklists: an "All Common" checklist, a "Laboratory General" checklist, and discipline checklists such as Chemistry, Hematology and Microbiology. Each requirement has an ID, a phase designation and "evidence of compliance" notes that tell you what the inspector will want to see.

The CAP Guide to Accreditation explains the program's expectations, and CAP's CAP Advance Document Review guidance describes options for sharing documents with the inspection team ahead of time. Requirements change with each checklist edition, so always work from the edition CAP has issued to your lab.

Start with the right checklists

  1. Download your customized checklists from CAP based on your activity menu.
  2. Assign an owner to each checklist or section.
  3. For each requirement, record where the evidence lives and when it was last reviewed.
  4. Pay attention to the "evidence of compliance" text; that is your document list.

survey preparation areas inspectors review closely

Document control

The CAP Guide to Accreditation describes a document control system in which every policy and procedure is current, authorized by the laboratory director before implementation, reviewed at least every two years by the director or designee, and in which discontinued procedures are retained for at least two years (five years for transfusion medicine). Staff must have read the procedures relevant to their work. See lab document control.

Personnel and competency

Personnel files must show qualifications for each role and competency assessments covering the six CLIA elements for each test system, completed semiannually during the first year and annually thereafter, consistent with 42 CFR 493.1451. Delegation of responsibilities by the director must be documented. See competency tracking.

Quality control and quality management

Inspectors look for QC run as required, reviewed by supervisory staff, with documented corrective action when rules are violated, plus a written quality management plan. Labs using an Individualized Quality Control Plan (IQCP) need the risk assessment, QC plan and quality assessment documents. See quality management software.

Proficiency testing

PT must be performed for each analyte as required, results reviewed by the director or designee, and unacceptable results investigated. Inter-laboratory communication and referral of PT samples are prohibited, as in 42 CFR 493.801. See PT tracking.

Method validation and verification

CAP's advance document review guidance lists method, test and instrument validation, laboratory-developed test documentation, IQCPs, the quality management plan and delegation of responsibility among the documents labs may upload ahead of the inspection. Have these complete and signed.

Equipment, calibration and environment

Maintenance, function checks, calibration and calibration verification, temperature monitoring for reagents and refrigerators, and pipette checks all need complete records with review.

A survey preparation timeline

WhenWhat to do
ContinuouslyRecord QC review, maintenance, competency and document reviews as they happen, not before the inspection.
Interim yearComplete the self-inspection CAP requires between on-site inspections, using the current checklists, and correct findings.
Before the window opensConfirm your document review option in CAP's Organization Profile; check that delegation, quality management plan, IQCP and validation documents are current.
OngoingRun mock tracers: pick a patient result and trace it back through specimen, QC, calibration, analyst competency and procedure version.

Common gaps to look for

  • Competency completed for some but not all test systems an employee performs.
  • Procedures past their two-year review, or bench copies that differ from the approved version.
  • QC rule violations without documented corrective action, or no evidence of supervisory review.
  • PT results not reviewed or signed, or unacceptable results without investigation.
  • Instrument comparison or calibration verification records missing for the period.

How myLabCompliance supports CAP survey preparation

myLabCompliance describes a CLIA, CAP and COLA survey-ready program across its plans. Its modules map to the areas above: personnel qualification and license tracking, competency assessment, document control with version control, review cycles and e-signatures, QC scheduling with Levey-Jennings charts and Westgard detection, QC investigation tracking, instrument maintenance and calibration, PT enrollment and corrective action, and validation studies. Lab Director and Technical Consultant dashboards help owners see open items. The software organizes evidence; your laboratory director and team remain responsible for compliance and accreditation outcomes.

Frequently asked questions

How often does CAP inspect laboratories?

CAP-accredited laboratories are inspected on a two-year cycle, with a self-inspection required in the intervening year. Check CAP's current program requirements for details.

What checklists does CAP use?

CAP uses the All Common and Laboratory General checklists plus discipline-specific checklists, customized to each laboratory's activity menu.

How often must procedures be reviewed under CAP?

CAP's Guide to Accreditation states that policies and procedures must be reviewed at least every two years by the laboratory director or designee.

Can documents be shared with CAP inspectors in advance?

Yes. CAP's advance document review guidance offers options including sharing during the inspection, guest access to a document management system, or uploading selected documents in Organization Profile.

Does myLabCompliance guarantee a successful CAP inspection?

No. It organizes the records and reviews your team performs. Accreditation decisions are made by CAP.

See myLabCompliance with your own lab's workflow

Book a demo and bring the compliance task that takes the most of your week. Product details are on mylabcompliance.io.

Book a demo

Sources: CAP Accreditation Checklists · CAP Guide to Accreditation · CAP Advance Document Review guidance · 42 CFR Part 493 (eCFR)

This page is general information about laboratory regulations, not legal or regulatory advice. Confirm current requirements with CMS, your state agency, and your accrediting organization.