How CLIA surveys work
CLIA is administered by CMS, with state survey agencies performing inspections for labs that are not accredited. Certificates are valid for two years, and labs holding a Certificate of Compliance are generally surveyed on that cycle. Accredited labs (Certificate of Accreditation) are inspected by their accreditor instead, though CMS may perform validation surveys. Surveyors follow the CMS State Operations Manual, Appendix C, which lists each regulation with a "D-tag" and interpretive guidance. The regulations themselves are in 42 CFR Part 493 (eCFR).
This checklist covers the most commonly reviewed areas for moderate and high complexity testing. It is not a substitute for the regulation text or your accreditor's checklist.
1. Certificate and administration
- CLIA certificate is current and matches the complexity of testing performed and the test menu.
- Changes in ownership, name, location or director were reported to CMS/state within the required time (42 CFR 493.51).
- The laboratory director's responsibilities are documented, including any delegation in writing (42 CFR 493.1445 for high complexity, 42 CFR 493.1407 for moderate).
2. Proficiency testing (Subpart H)
- Enrolled in a CMS-approved PT program for every regulated analyte, specialty and subspecialty tested (42 CFR 493.801).
- PT samples tested with patient workload by personnel who routinely test patients, using routine methods.
- No inter-laboratory communication about PT results and no referral of PT samples to another lab before the submission deadline (42 CFR 493.801(b)).
- PT records, including attestations signed by the analyst and director, retained (42 CFR 493.801(b)(5)).
- For tests without required PT, accuracy verified at least twice a year (42 CFR 493.1236(c)).
- Unsatisfactory results investigated with documented corrective action. More detail: proficiency testing tracking.
3. General laboratory systems and preanalytic
- Confidentiality of patient information, specimen identification and integrity (42 CFR 493.1231, 42 CFR 493.1232).
- Complaint investigations and communications documented (42 CFR 493.1233, 42 CFR 493.1234).
- Written test requests and specimen submission, handling and referral procedures (42 CFR 493.1241, 42 CFR 493.1242).
4. Analytic systems and quality control
- Procedure manual complete, approved, signed and dated by the director; changes approved before use (42 CFR 493.1251). See document control.
- Reagents and materials labeled, stored per manufacturer and not used past expiration (42 CFR 493.1252).
- Performance specifications verified for each unmodified FDA-cleared test before reporting patient results: accuracy, precision, reportable range and reference intervals (42 CFR 493.1253). Modified or lab-developed tests require establishment of additional specifications.
- Maintenance and function checks performed and documented as required by the manufacturer (42 CFR 493.1254).
- Calibration and calibration verification performed at least every six months and when specified events occur (42 CFR 493.1255).
- Control procedures meet 42 CFR 493.1256, including at least two levels of control each day of testing for quantitative tests, unless an IQCP is in place as described in the SOM.
- Comparison of test results when multiple methods or instruments are used for the same test, twice a year (42 CFR 493.1281).
- Corrective actions documented when QC, calibration or other checks fail (42 CFR 493.1282). See QC and quality management.
- Test records retained per 42 CFR 493.1283 and 42 CFR 493.1105.
5. Postanalytic systems
- Reports include required elements and are sent promptly to the authorized person; critical values communicated (42 CFR 493.1291).
- Corrected reports issued and documented when errors are found.
6. Personnel (Subpart M)
- Director, technical consultant or technical supervisor, clinical consultant, general supervisor (high complexity) and testing personnel meet qualifications, with diplomas, transcripts and licenses on file (42 CFR 493.1405–42 CFR 493.1495).
- Competency assessed for each person using the six required procedures, at six months and one year for new testing personnel and annually thereafter (42 CFR 493.1413, 42 CFR 493.1451). See competency tracking.
- Training documented before patient testing.
7. Quality assessment
- Ongoing quality assessment of preanalytic, analytic and postanalytic systems, with problems identified, corrected and reviewed with staff (42 CFR 493.1239, 42 CFR 493.1249, 42 CFR 493.1289, 42 CFR 493.1299).
How myLabCompliance helps with this checklist
myLabCompliance gives each of these areas a home: personnel qualification tracking with CLIA pathways and expiration alerts, competency assessments, document control with version history and e-signatures, automated QC scheduling with Levey-Jennings charts and Westgard rule detection, instrument maintenance and calibration, PT tracking and a lab calendar. Lab Director and Technical Consultant dashboards show what is open. If you also hold accreditation, see the CAP and COLA guides.
Frequently asked questions
How often are CLIA labs inspected?
CLIA certificates are valid for two years, and labs with a Certificate of Compliance are generally surveyed by the state agency on that cycle. Accredited labs are inspected by their accreditor.
What is a D-tag?
D-tags are the identifiers used in CMS State Operations Manual Appendix C to reference specific CLIA regulations and their interpretive guidance. Deficiencies on a CMS-2567 cite them.
How often must calibration verification be done?
At least once every six months and whenever certain events occur, such as a major reagent change or major maintenance, under 42 CFR 493.1255, unless the manufacturer's procedure meets the stated criteria.
How long do CLIA records have to be kept?
Generally at least two years under 42 CFR 493.1105, with longer periods for some records such as immunohematology and pathology.
Can software do the checklist for me?
No, but tools like myLabCompliance keep the evidence for each item organized and flag due dates so gaps surface before a survey.
See myLabCompliance with your own lab's workflow
Book a demo and bring the compliance task that takes the most of your week. Product details are on mylabcompliance.io.
Book a demoSources: 42 CFR Part 493 (eCFR) · CMS State Operations Manual, Appendix C · CMS CLIA program
This page is general information about laboratory regulations, not legal or regulatory advice. Confirm current requirements with CMS, your state agency, and your accrediting organization.